Health Canada Recall: Life-Saving Medical Equipment Recalled due to Misassembled Issue

Health Canada Recall Life-Saving Medical Equipment Recalled due to Misassembled Issue

In a critical safety notice issued by Health Canada, healthcare providers are being warned about a serious recall affecting Adult Manual Resuscitators due to a misassembly issue that could lead to delayed resuscitation and life-threatening outcomes, including hypoxia, hypercapnia, organ failure, and even death.


Product Overview

Recalled Device:
Adult Manual Resuscitator with mask, bag reservoir, and filter

Manufacturer:
Medline (via contract manufacturer)

Model/Catalog Number:
CPRM1116F

Recall Start Date:
May 9, 2025

Pain Relief Stick Recalled in Canada due to undeclared ingredients and missing risk

Popular Dessert Recalled in Canada Over Dangerous E. Coli Contamination — Check Your Freezer Now

Nationwide Food Recall: Cucina & Amore Pesto Pasta Sauce Pulled from Canadian Shelves Due to Undeclared Allergens

Multiple Foods Recalled Across Canada Due to Listeria, Salmonella, Insects, and Wood Contamination

Health Canada Issues Urgent Recall of $130 Face Cream Sold at Shoppers Drug Mart


What’s the Problem?

The issue stems from a critical misassembly of the resuscitator components. According to Medline, the filter that should have been pre-attached at the exhalation port was instead connected to the patient port, where the mask is typically attached.

This means that if the device is used without detecting the misplacement, it could cause:

  • Obstructed airflow
  • Delay in oxygen delivery
  • Potential failure to properly ventilate the patient

Such a failure during an emergency could lead to:

  • Hypoxia (oxygen deficiency in the body)
  • Hypercapnia (excess carbon dioxide)
  • Organ damage or failure
  • Death, particularly in critical care or emergency resuscitation scenarios

Affected Lot Numbers

The following serial or lot numbers have been identified in the recall:

526796, 526804, 526797, 526805, 526817, 526782, 526802, 526815, 526807, 526811, 526800, 526806, 526809, 526814, 526816, 526798, 526808, 526790, 526810, 526818

Healthcare providers and institutions are advised to immediately inspect their inventory and quarantine any affected units.


What Healthcare Providers Should Do

  1. Stop using the affected product immediately.
  2. Check the model number (CPRM1116F) and lot/serial numbers listed above.
  3. If you find any affected resuscitators:
    • Do not use them.
    • Contact Medline for further instructions and replacement procedures.
  4. Report any incidents or adverse health events to Health Canada’s MedEffect system.

Why This Matters

Manual resuscitators are life-saving emergency devices, commonly used in:

  • Emergency departments
  • ICUs
  • EMS and paramedic response
  • Surgical recovery units

Any malfunction or delay during resuscitation could mean the difference between life and death, particularly in time-sensitive scenarios involving cardiac arrest, respiratory failure, or trauma.


Final Reminder

If you’re a hospital administrator, paramedic, clinician, or involved in procurement of medical supplies, act immediately to:

  • Verify your stock
  • Prevent use of affected devices
  • Ensure patient safety

Your prompt response could prevent a potentially tragic incident.